FLUKE 1524 HANDHELD THERMOMETER READOUT

Availability : in stock
Share:

Key features

  • Measure, graph, and record PRTs, thermocouples, and thermistors, supporting a wide range of temperature sensors.
  • Accuracy up to ±0.002 °C.
  • Quick sample readings in as little as 0.3 seconds.
  • 1524 has two channels for simultaneous measurements, doubling productivity.

Specifications: Fluke Calibration 1524 Handheld Thermometer Readout

Specifications
Input channels 2
Logging 15,000 time and date stamped; 25 readings with statistics
Sample interval
(normal)
1 second
Sample interval (fast mode) 0.3 seconds (see technical manual for details)
Sensor types PRTs, RTDs, thermistors, and thermocouples
Thermocouple types C, E, J, K, L, M, N, T, U, B, R, S
Operating temperature -10°C to 60°C (Best accuracy 13°C to 33°C)
Power requirements 3 AA alkaline batteries
Size 96 x 200 x 47 mm
(3.75 x 7.9 x 1.86 in)
Weight 0.65 kg (1.4 lb)
Environmental conditions for best accuracy 13°C to 33°C
Millivolt range and accuracy –10 mV to 75 mV ±(0.005% + 5 μV)
Resistance range and accuracy 0 Ω to 400 Ω ±(0.004% + 0.002 Ω)
200 Ω to 50 kΩ ±(0.01% + 0.5 Ω)
50 kΩ to 500 kΩ ± (0.03%)
Temperature coefficient, voltage (–10°C to 13°C, +33°C to 60°C) ±(0.001%/°C + 1 μV/°C)
Temperature coefficient, resistance (–10°C to 13°C, +33°C to 60°C) 0.0008%/°C + 0.0004 Ω (0 Ω to 400 Ω)
0.002%/°C + 0.1 Ω (0 Ω to 50 kΩ)
0.06%/°C + 0.1 Ω (50 kΩ to 500 kΩ)
Excitation current, resistance 1 mA (0 Ω to 400 Ω)
10 μA (0 Ω to 50 kΩ)
2 µA (50 kΩ to 500 kΩ)
Thermocouple Equivalent Temperature Accuracies (Readout Only)
Type B ±0.85°C from 600°C to 800°C
±0.68°C from 800°C to 1000°C
±0.57°C from 1000°C to 1800°C
Type C ±0.32°C from 100ºC to 550°C
±0.71°C from 550°C to 2300°C
Type E ±0.52°C from -200°C to 0°C
±0.22°C from 0°C to 950°C
Type J ±0.52°C from -200°C to 0°C
±0.23°C from 0°C to 1200°C
Type K ±0.61°C from -200°C to 0°C
±0.24°C from 0°C to 1370°C
Type L ±0.36°C from -200°C to 0°C
±0.23°C from 0°C to 1370°C
Type M ±0.26°C from -20°C to 0°C
±0.25°C from 0°C to 400°C
±0.22°C from 400°C to 1400°C
Type N ±0.72°C from -200°C to 0°C
±0.28°C from 0°C to 1300°C
Type R ±1.09°C from -20°C to 0°C
±0.97°C from 0°C to 500°C
±0.49°C from 500°C to 1750°C
Type S ±1.05°C from -20°C to 0°C
±0.95°C from 0°C to 500°C
±0.56°C from 500°C to 1750°C
Type T ±0.60°C from -200°C to 0°C
±0.25°C from 0°C to 400°C
Type U ±0.54°C from -200°C to 0°C
±0.24°C from 0°C to 400°C
Accuracies are based on internal reference junction compensation. Refer to technical manual for equivalent accuracies with an external reference junction.
1524 Accuracies with Selected Probes (±°C)
-200°C 5616-12: 0.014
5615-6: 0.025
5627A-12: 0.027
5610-9: n/a
0°C 5616-12: 0.021
5615-6: 0.021
5627A-12: 0.049
5610-9: 0.009
100°C 5616-12: 0.027
5615-6: 0.028
5627A-12: 0.065
5610-9: 0.009
300°C 5616-12: 0.040
5615-6: 0.043
5627A-12: 0.103
5610-9: n/a
420°C 5616-12: 0.050
5615-6: n/a
5627A-12: 0.130
5610-9: n/a
Includes readout accuracy, probe calibration, and probe drift
PRT Equivalent Temperature Accuracy (Readout only)
–100°C ±0.011
0°C ±0.015
100°C ±0.019
200°C ±0.023
400°C ±0.031
600°C ±0.039
Thermistor Equivalent Temperature Accuracy (Readout Only)
0°C ±0.002
25°C ±0.003
50°C ±0.006
75°C ±0.014
100°C ±0.030